13485cert

Archive for the ‘Class III’ Category

Who’s Afraid of the Proposed European Scrutiny Process?

In CE Mark, CE Medical, Class IIb, Class III, Medical CE, PMA, Scrutiny Process on October 12, 2012 at 2:22 am

For those of you that are not familiar with the “Scrutiny Process”, I am referring specifically to Article 44 of the proposed EU regulations for medical devices. This process is first alluded to at the end of section 3.5 in the “Explanatory Memorandum” (i.e. – the 13 pages preceding the proposal for the regulation of medical devices).

I was looking for a video that matched up with my title and when I saw this TECHNO music video.

The US already has a pre-market approval process that we fondly refer to as the PMA process. In response to the PIP scandal, the European Parliament’s ENVI Committee (Committee on the Environment, Public Health and Food Safety) proposed a pre-market approval process as part of a press release issued on April 25, 2012. In response to this political pressure, the Commission has proposed a “Scrutiny Process” that involves preparation of a Notified Body “Summary Evaluation Report” and verification that the conformity assessment was adequate by the Coordinating Competent Authority. A similar process is outlined in MEDEV 2.11/1 rev. 2, a guidance document regarding animal tissues, and the Commission Regulation (EU) No 722/2012 of 8 August 2012. The proposed scrutiny process allows competent authorities to take a “second look” and review the findings of the Notified Body that would be issuing a CE Certificate for these high risk devices. The review process is supposed to be concluded within 60 days, but the review time limit is suspended if the Competent Authorities request additional information or product samples within the first 30 days.

In section 3.5 of the Explanatory Memorandum, the Commission states that this scrutiny process “should be the exception rather than the rule and should follow clear and transparent criteria.” The criteria for invoking the scrutiny process are defined in five points 5a) through 5e) of Article 44. The five points leave room for interpretation by Competent Authorities, and the medical device industry is concerned that the review process for Class IIb and Class III devices will be delayed by at least 60 days on a regular basis. The process could easily be delayed by as much as six months when there are requests for additional information and samples.

The “Legislative Financial Statement” (i.e. – the 19 pages immediately following the proposal for the regulation of medical devices) defines a monitoring process for the scrutiny process in the “Indicator of results and impact” (Section 1.4.4). The risk of delaying access to market for innovative devices is also identified in the “Risk(s) identified” (Section 2.2.1). Therefore, the need for a control mechanism is identified in “Control method(s) envisaged” (Section 2.2.2). This will be the responsibility of the Commission to draft a guidance document to define the control method(s). Until industry has an opportunity to review such a guidance document, executives will continue to voice their concerns and apply their own political pressure to the European Parliament.

Advertisement

What is the Design Input?

In 510(k), CE Mark, CE Medical, Class IIb, Class III, Design & Development, Design Inputs, Design Outputs, Design Validation, Design Verification, ISO 13485, Medical CE, Medical Device, Risk Management on May 12, 2012 at 7:09 pm

Micky, this is for you.

I have been directly involved in dozens of design projects throughout my career, and during the past three years I have audited 50+ Design Dossiers for CE Marking of Medical Devices. Throughout most of these design projects, I have noticed one common thread—a misunderstanding of design inputs.

ISO 13485 identifies the requirements for Design Inputs. These requirements are:

  1. Functional (7.3.2a)
  2. Performance (7.3.2a)
  3. Safety (7.3.2a)
  4. Statutory / Regulatory (7.3.2b)
  5. Previous and Similar Designs (7.3.2c)
  6. Essential Requirements (7.3.2d)
  7. Outputs of Risk Management (7.3.2e)
  8. Customer Requirements (7.2.1)
  9. Organizational Requirements (7.2.1)

The most common error seems to be the failure to include the outputs of risk management. For those of you that have used design FMEA’s—that’s what the right-hand columns are for. When you identify suggested actions to mitigate risks with the current design, these actions should be translated into inputs for the “new and improved” model.

The second most common error seems to be failure to consider regulatory requirements. There are actually two ways this mistake is frequently made: 1) Canadian MDR’s were not considered as design inputs for a device intended for Canadian medical device licensing, and 2) an applicable ISO Standard was not considered (i.e. – “State of the Art” is Essential Requirement 2 of the Medical Device Directive or MDD).

The third most common error, and the one that drives me crazy, is confusion of design outputs and design inputs. For example: an outer diameter of 2.3 +/- 0.05 mm is not a design input for a 7 French arterial catheter. This is a design output. The user need might be that the catheter must be small enough to fit inside the femoral artery and allow interventional radiologists to navigate to a specific location to administer therapy. Validation that the new design can do this is relatively straight forward to evaluate in a pre-clinical animal model or a clinical study. The question is, “What is the design input?”
Design inputs are supposed to be objective criteria for verification that the design outputs are adequate. One example of a design input is that the catheter outer diameter must be no larger than a previous design that is an 8 French catheter. Another possible design input is that the catheter outer diameter must be less than a competitor product. In both examples, a simple measurement of the OD is all that is required to complete the verification. This also gives a design team much more freedom to develop novel products than a narrow specification of 23 +/- 0.05 mm allows for.

If you are developing a Class II medical device for a 510(k) submission to the FDA, special controls guidance documents will include design inputs. If you are developing a Class IIa, Class IIb or Class III medical device for CE marking, there is probably an ISO Standard that lists functional, performance and safety requirements for the device. Regulatory guidance documents and ISO Standards usually reference test methods and indicate acceptance criteria. When you have a test method and acceptance criteria defined, it is easier to write a verification protocol. Therefore, design teams should always strive to document design inputs that reference a test method and acceptance criteria. If this is not done, verification protocols are much more difficult to write.

In my earlier example, the outer diameter of 2.3 +/- 0.05 mm is a specification. Unfortunately, many companies would document this as an input and use the final drawing as the output. By making this mistake, “verification” is simply to measure the outer diameter to verify that it matches the drawing. This adds no value and if the specifications are incorrect the design team will not know about it.

A true verification would include a protocol that identifies the “worst-case scenario” and verifies that this still meets the design input requirements. Therefore, if the drawing indicates a dimensional tolerance of 2.3 +/- 0.05, “worst-case” is 2.35 mm. The verification process is to measure either a previous version of the product or a competitor’s catheter. The smallest previous version or competitor catheter tested must be larger than the upper limit of the design output for outer diameter of the new catheter.

How do you control design changes?

In Change Control, Class IIb, Class III, Design & Development, ISO, ISO 13485, ISO 14971, Medical Device, PMA, Quality, Quality Management Systems, Risk Management on May 4, 2012 at 4:59 am

Of JB’s recommended artists, the Josh Abbott Band was probably my favorite. I especially liked this one. I hope every man is lucky enough to know a girl like Texas. I’m lucky enough to have married a girl that grew up in Texas. They are something special.

We have been discussing the best ways to control design changes at work, and I thought it might present an opportunity to have more of an interactive discussion with my readers.

During my rounds as a 3rd party auditor, I have seen quite a few design control procedures. The most complex consisted of 19 procedures (NOT recommended, but there were no nonconformities). The most simple consisted of one 4-page procedure, which I wrote, but I would never recommend being this brief. I have created a couple of polls in my LinkedIn profile for you to respond to if you would like to share your own company’s “design control stats”:

http://linkd.in/IJtoBL

The problem I see is that most projects are not new product designs. Sometimes the projects are not even major design changes. I think most changes involve supplier changes, component specification improvements, and design for manufacturability. These changes require review and approval of changes. These changes must also be recorded and retained as a Quality Record.

My own personal preference is to always open a design project—no matter how small the change is. In order to make the process flexible, I also prefer to define how many design reviews each project will have in the design plan rather than mandating that design reviews be held in a stage-gate fashion for 100% of projects.

Most companies will have a table of requirements with columns added to indicate if the requirements are mandatory for the project or optional. For example, “risk management plan needs to be updated? Yes/No.” I like this approach, because the table of requirements makes the decision making systematic.

Sometimes a change is only to a work instruction for a step in the manufacturing process. In these cases, some companies will use a document change order process to supplement the engineering change order process.

My feeling is that more complex products (i.e. – Class IIb & Class III in EU and Class III/PMA in US) will require more stringent design controls for the change. What does your company do to control design changes?

%d bloggers like this: